PCR Workstations

Get Results—Not Background!
AirClean® Systems PCR workstations allow you to maintain a clean work area during PCR preparation. The AC600LFUV has a built-in 254 nm ultraviolet light that allows the operator to irradiate tubes, flasks and pipettes between amplifications. The 254 nm ultraviolet light will effectively destroy potentially contaminating DNA and RNA within minutes.
The workstations feature a unique interlocking sash with a safety overlap to prevent gaps. The sash also has a safety switch that automatically turns off the ultraviolet light and activates the blower and fluorescent light when the sash is raised. This revolutionary uni-body design provides unmatched user protection from harmful ultraviolet rays during the sterilization process. HEPA filtered units have a built-in monitoring system to alert the user when to change the HEPA filter.
Typical Workstation Airflow:
- Room air enters at "A"
- Air forced through HEPA filter creating laminar flow
- Clean laminar flow air enters at "B" and exits at "C"
- UV light sterilizes pipette tips, tubes and flasks
Features:
- Class 100 vertical laminar flow
- 254 nm ultraviolet light with 0-15 minute automatic timer
- Bi-fold sash with safety interlock
- 360-degree visibility
- Integral polypropylene base for easy cleaning
- Low-profile design using all thermoplastic construction
- Built-in fluorescent light
- HEPA filter monitor automatically indicates when filter change is required
- Available in 110V or 220V models
Options:
- ACA1037 Two Speed Blower—increases filter life and enables ultra-quiet background operation.
- ACA1038 Variable Blower Controller—for 0-100% control of blower speed.
- ACA1011 Custom Sturdy Cart for AC632—allows unit to be portable.
- ACA1039 Custom Sturdy Cart for AC648—allows unit to be portable.
- ACA1040 Removable Shelf—allows pipette tips and tubes to be placed closer to UV light for irradiation. Note: Two shelves required for AC648.
- ACA1041 Decontamination Spray—Formulated to provided high-level disinfection without damaging workstation components.
| CAT. # | DESCRIPTION |
| AC632DB | Dead Air Box, 32" wide |
| AC632LF | Class 100 laminar flow, 32" wide |
| AC632LFUV | Combination class 100 Laminar flow and UV light sterilization, 32" wide |
| AC648DB | Dead Air Box, 48" wide |
| AC648LF | Class 100 laminar flow, 48" wide |
| AC648LFUV | Combination Class 100 Laminar flow and UV light sterilization, 48" wide |
AC600 Series - Three Models:
DB – Circulation free environment with UV light sterilization. This unit has a UV light with 0-15 minute timer for sterilization between amplifications. A safety interlock switch prevents exposure to UV light when sash is raised.
LF – Provides Class 100 vertical laminar flow air through a 99.997% HEPA filter.
LFUV – Combines a Class 100 clean air environment with UV light sterilization for optimal protection from sample contamination. Raising the bi-fold sash turns off the UV light and enables the blower.
| AirClean® AC600 Series Specifications | ||||||
| AC632DB | AC632LF | AC632LF | AC648DB | AC648LF | AC648LFUV | |
| Width | 32" | 32" | 32" | 48" | 48" | 48" |
| Depth | 24" | 24" | 24" | 26" | 26" | 26" |
| Height | 30" | 30" | 30" | 32" | 32" | 32" |
| Ship Weight | 100 lbs. | 125 lbs. | 150 lbs. | 125 lbs. | 150 lbs. | 200 lbs. |
| Noise Level | N/A | <49DBA | <49DBA | N/A | <53DBA | <53DBA |
| Voltage | 110V or 220V | 110V or 220V | 110V or 220V | 110V or 220V | 110V or 220V | 110V or 220V |
| Power | 20 Watts | 200 Watts | 200 Watts | 40 Watts | 400 Watts | 400 Watts |
| UV Light | 15W 254 nm | N/A | 15W 254 nm | 2-15W 254 nm | N/A | 2-15W 254 nm |
| Blower | N/A | Sparkless/Brushless | Sparkless/Brushless | N/A | Sparkless/Brushless | Sparkless/Brushless |
| UV Light Timer | 0-15 minutes | N/A | 0-15 minutes | 0-15 minutes | N/A | 0-15 minutes |
| Safety Interlock | Yes | N/A | Yes | Yes | N/A | Yes |
| Air Velocity | N/A | 100LFM/60LFM | 100LFM/60LFM | N/A | 100LFM/60LFM | 100LFM/60LFM |